Glove Box Regeneration Steps

Uit Schoolwerkplan SBS Munsterbilzen - Hoelbeek - Martenslinde
Naar navigatie springen Naar zoeken springen

one) Regeneration gasoline: a combination of hydrogen and inert fuel (hydrogen accounts for 5% -10%)

2) Amount: glove box manufacturers 40L fuel cylinder 10Mpa

three) Regenerative fuel strain: .05Mpa (the gas use per fifty percent hour in the state of .05Mpa is that the cylinder strain drops by one.5Mpa. If it exceeds 1.5Mpa, the strain of the strain regulator’s timetable will be minimized, if it is fewer than one.5Mpa Improve the pressure of the stress regulator timetable)

four) Approach: 3 hrs heating-3 hours heating and air flow-three hrs vacuuming-1 hour pumping-ten hrs cooling

five) Techniques:

① Join the regeneration gas (verify the pipeline for leaks immediately after link: right after connecting, open up the principal valve of the pressure minimizing valve to the utmost, open up the auxiliary meter to .1Mpa, flip off the most important valve, wait around fifteen minutes, if the pressure of the auxiliary meter continues to be at .one Mpa remains unchanged, indicating no air leakage

>②Set the pressure of the auxiliary gauge of the tension lessening valve to .1Mpa (the pressure for the duration of air flow is about .06-.08MPa

>③ Near the cycle and analyz

r>④Turn on the vacuum pu

r>⑤ Begin regenerat

br>⑥ The new plan finishes at 20h (thoroughly clean ahead of cleansing, water and oxygen index <200p


6) Note: Do not cut off the power and gas during the regeneration process (especially the first 6 hours)! Some pressure reducing valves are not allowed, and the pressure needs to be adjusted appropriately to ensure continuous ventilation for 3 ho


7) Observation: whether the regeneration gas flowmeter shows about 20L / min, otherwise adjust the pressure of the pressure reducing valve.